It uses ultrasonic technology to measure the air velocity. The detection head consists of a single transmitter and a row of two receivers. When the wind direction is the same as the arrow direction, the transmission time of A will increase with the increase of wind speed, and the transmission time of B will increase with the increase of wind speed. The wind speed increases and decreases. The wind direction is opposite to the arrow direction. The transmission time of A decreases and the transmission time of B increases. According to the phase difference of the ultrasonic signals received by the two receivers, the wind speed and direction can be calculated.
Technical Index:
Measuring Range: -30m/s~+30m/s, the highest measuring wind speed is 80m/s (the measuring range can be adjusted)
Average Time: 0 seconds to 60 minutes
Measurement Accuracy: ≤±0.1m/sec
Analog Output: 2*4/2/0~20mA current isolation output maximum load impedance 1000Ω
Switch Output: 2*SPCO passive relay contacts, 0.25A/125VAC, 1A/48VDC, 0.25A/100VDC, 1A/30VDC configured as fault alarm or limit value alarm
Communication Interface: RS232/RS485/RS422
Shell Protection Level: plastic-sprayed aluminum shell, IP68
Weight: 5.5kg,
Power Supply: 76~265VAC
Working Temperature: -50~75℃